首页 | 本学科首页   官方微博 | 高级检索  
     


Post-injection hybridization of complementary DNA strands on capillary electrophoresis platforms: A novel solution for dsDNA artifacts
Authors:Robert S. McLaren   Martin G. Ensenberger   Bruce Budowle   Dawn Rabbach   Patricia M. Fulmer   Cindy J. Sprecher   Joseph Bessetti   Terri M. Sundquist  Douglas R. Storts
Affiliation:aPromega Corporation, 2800 Woods Hollow Road, Madison, WI 53711, United States;bFederal Bureau of Investigation, 2501 Investigation Parkway, Quantico, VA 22135, United States
Abstract:
Several laboratories have reported the occurrence of a split or n − 1 peak at the vWA locus in PowerPlex® 16 and PowerPlex® ES amplification products separated on 4- and 16-capillary electrophoresis instruments. The root cause of this artifact is post-PCR reannealing of the unlabeled, unincorporated vWA primer to the 3′-end of the tetramethylrhodamine (TMR)-labeled strand of the vWA amplicon. This reannealing occurs in the capillary post-electrokinetic injection. The split peak is eliminated by incorporation into the loading cocktail of a sacrificial hybridization sequence (SHS) oligonucleotide that is complementary to the vWA primer. The SHS preferentially anneals to the primer instead of the TMR-labeled strand of the vWA amplicon. In addition, the n − 10/n − 18 artifact that may be seen at the vWA locus was determined to be due to double-stranded amplicon formed post-electrokinetic injection into the capillary. This was also eliminated by adding in two Complementary Oligo Targets (COT1 and COT2) in addition to the SHS oligonucleotide into the loading cocktail. These three oligonucleotides are complementary to the 33 bases at the 5′-end of the unlabeled vWA amplicon strand and the 60 bases at its 3′-end and therefore compete for hybridization to the TMR-labeled amplicon strand. Incorporation of these three oligonucleotides in the Internal Lane Standard 600 (ILS600) eliminate both the split peak and n − 10/n − 18 artifact in PowerPlex® 16 and PowerPlex® ES amplification products without affecting sizing of alleles at the vWA locus or any locus in the PowerPlex® 16, PowerPlex® Y, PowerPlex® ES, AmpFlSTR® Profiler Plus® ID, AmpFlSTR® Cofiler®, and AmpFlSTR® SGM Plus® kits.
Keywords:Split peak   Secondary structure   Denaturation   vWA   Artifacts   Short tandem repeat (STR)   PowerPlex®     Capillary electrophoresis
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏