首页 | 本学科首页   官方微博 | 高级检索  
     


Validation of Reduced Reagent Volumes in the Implementation of the Quantifiler® Trio Quantification Kit
Authors:Yoonjung Cho Ph.D.  Hyo Sook Kim M.S.  Min‐Hee Kim M.S.  Moonhee Park M.S.  Hansol Kwon M.S.  Yang Han Lee Ph.D.  Dong Sub Lee M.S.
Affiliation:Forensic DNA Division, National Forensic Service, Wonju, Gangwon‐do, Korea
Abstract:The Quantifiler® Trio Quantification Kit has been developed to quantify the total amount of amplifiable and human male DNA in samples and to estimate the extent of DNA degradation. To minimize the cost of DNA quantification, we evaluated kit performance using a reduced volume of reagents (1/10‐volume) using DNA samples of varying types and concentrations. Our results demonstrated concordance between the manufacturer's method and the low‐volume method for DNA quantification, DNA degradation index estimation, and human male DNA quantification. We confirmed the practical utility of the low‐volume method with 109 casework samples by evaluating short tandem repeat (STR) profiling success with respect to DNA quantity and quality. We also defined a cutoff value for DNA quantity to ensure reliable STR results. Using a reduced volume of reagents, 10 times more reactions per kit are possible; accordingly, this method reduces the cost of DNA quantification, while maintaining performance.
Keywords:forensic science  Quantifiler®   Trio  human male DNA  degradation index  DNA quantification  short tandem repeat
设为首页 | 免责声明 | 关于勤云 | 加入收藏