首页 | 本学科首页   官方微博 | 高级检索  
     


Discrepancies between forensic identification kits explained by a laser power supply shutdown
Authors:Siméon Sylvaine  Puech Catherine  de Mazancourt Philippe
Affiliation:Laboratory of Biochemistry and Molecular Biology, UVSQ EA 2493, Hospital R. Poincare, 104 Bd R. Poincare, 92380 Garches, France.
Abstract:The occurrence of two null alleles for the same individual was suspected because of the discrepancy at two loci (vWA and THO1) between genotypes characterized by the mean of two different commercial kits. Reanalysis of the samples indicated there was no null allele, and that this abnormality was indeed related to an automatic laser power shutdown. The laser power supply was set to a value (6.4V) too close to the automatic shutdown threshold (6.0V). The alleles were undetected because of the absence of emission at the time of migration of the two alleles through the detection device. However, since no alarm is triggered in such an event, the users should monitor their laser power supply values.
Keywords:
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号