Evaluation of the Cozart RapiScan drug test system for opiates and cocaine in oral fluid |
| |
Authors: | Gail Cooper Lisa Wilson Claire Reid Lianne Main Chris Hand |
| |
Affiliation: | Cozart Bioscience Ltd., Abingdon, Oxfordshire, OX14 4RY, UK |
| |
Abstract: | The purpose of this study was to evaluate the efficiency of the Cozart® RapiScan (CRS) drug test system for detecting opiates and cocaine in oral fluid. Oral fluid samples were collected using the Cozart® RapiScan collection system from 358 donors who were receiving treatment for their addiction and were monitored for drug misuse. A further 103 oral fluid samples were collected from volunteer donors who were not drug users. The samples were analyzed in the laboratory using the two-panel Cozart® RapiScan cartridge for opiates and cocaine and confirmed using gas chromatography–mass spectrometry (GC–MS). The samples were stored frozen at −20 °C until analysis by GC–MS. The overall accuracy of the CRS for both opiates and cocaine was 100%. Samples spiked at 50% above and below the cut-off consistently gave negative and positive results respectively. A total of 88 samples were positive for various opiates and 111 samples were positive for cocaine and/or its metabolites. The CRS for opiates and cocaine in oral fluid, using a cut-off of 30 ng/mL morphine or benzoylecgonine equivalents in neat oral fluid, had overall efficiencies of 98% and 99%, respectively, versus GC–MS. A series of potential adulterants of oral fluid were evaluated and shown not to alter the outcome of the test result. |
| |
Keywords: | Oral fluid Cocaine Opiates On-site test Saliva |