A comparison of three automated DNA purification methods in Forensic casework |
| |
Authors: | Christina Valgren Sara Wester Oskar Hansson |
| |
Affiliation: | Swedish National Laboratory of Forensic Science, SE-58194 Linköping, Sweden |
| |
Abstract: | Manual Chelex®-100 and organic extractions (phenol/chloroform) are used as routine methods at the Swedish National Laboratory of Forensic Science, SKL. The aim of this study was to find an automated DNA purification system to replace the organic method. The following methods were evaluated and compared to each other and to the organic method used routinely; BioRobot® EZ1 with EZ1 DNA Investigator Kit and Card (Qiagen), iPrep™ Purification Instrument with iPrep™ ChargeSwitch® Forensic Kit and Card (Invitrogen), Magnatrix™ 1200 Workstation with the Magnatrix™ gDNA Blood Kit Forensic and two different protocols; Forensic protocol A and B (Magnetic Biosolutions). Blood on fats, cotton swabs, moist snuff, paper towels and leather, post-mortem blood and muscle tissue were extracted with the different methods. DNA concentration and quality of the electropherograms were examined. Individual comparisons between the four extraction methods showed that iPrep™ and Magnatrix™ 1200 gave significantly lower mean quantities compared to BioRobot® EZ1 and the organic extraction method (p < 0.05). There were no significant differences between the latter two. BioRobot® EZ1 generated the best results and is in the process of being validated for routine analysis at SKL. |
| |
Keywords: | Organic extraction BioRobot® EZ1 iPrep&trade Magnatrix&trade Magnetic bead Magtration® technology |