Developmental Validation of the Quantifiler® Duo DNA Quantification Kit for Simultaneous Quantification of Total Human and Human Male DNA and Detection of PCR Inhibitors in Biological Samples* |
| |
Authors: | Maura Barbisin Ph.D. Rixun Fang Ph.D. Cristin E. O’Shea B.S. Lisa M. Calandro M.P.H. Manohar R. Furtado Ph.D. Jaiprakash G. Shewale Ph.D. |
| |
Affiliation: | Applied Biosystems, 850 Lincoln Centre Drive, Foster City, CA 94404. |
| |
Abstract: | Abstract: The Quantifiler® Duo DNA Quantification kit enables simultaneous quantification of human DNA and human male DNA as well as detection of inhibitors of PCR in a single real-time PCR well. Pooled human male genomic DNA is used to generate standard curves for both human (ribonuclease P RNA component H1) and human male (sex determining region Y) specific targets. A shift in the cycle threshold (CT) values for the internal positive control monitors the presence of PCR inhibitors in a sample. The assay is human specific and exhibits a high dynamic range from 0.023 to 50 ng/μL. In addition, the multiplex assay can detect as little as 25 pg/μL of human male DNA in the presence of a 1000-fold excess of human female DNA. The multiplex assay provides assessment of the DNA extract and guidance for the selection of the appropriate AmpFℓSTR® Amplification Kit to obtain interpretable short tandem repeat profiles. |
| |
Keywords: | forensic science DNA quantification real-time PCR DNA analysis ribonuclease P RNA component H1 sex determining region Y human DNA human male DNA DNA typing |
|