Operator Exposure to Scatter Radiation from a Portable Hand-held Dental Radiation Emitting Device (Aribex™ NOMAD™) While Making 915 Intraoral Dental Radiographs*
Authors:Robert A Danforth DDS  Edward E Herschaft DDS  MA  John A Leonowich PhD
Institution:1. Associate Professor, Clinical Sciences (Oral and Maxillofacial Radiology), UNLV School of Dental Medicine, Las Vegas, NV.;2. Professor, Biomedical Sciences (Oral and Maxillofacial Pathology), UNLV School of Dental Medicine, Las Vegas, NV.;3. Radiation Safety Officer, University of Nevada—Las Vegas, NV.
Abstract:Abstract: Operator exposure to backscatter radiation while using an Aribex™ NOMAD™ radiation emitting device (a portable, self-contained, cordless, hand-held dental X-ray unit) was determined while the operator employed various typical and atypical use scenarios during the exposure of 715 digital and/or film-based dental radiographs and 200 study control exposures. Study data was compared to the radiation safety occupational exposure annual Maximum Permissive Dose (MPD) of 50 mSv (5000 mrem) to determine the possible exposure risk to an unprotected operator using this device. The results showed the reproductive organs received the highest dose and the thyroid the least. The average operator whole body dose for the study was determined to be 0.047 mSv (4.47 mrem) or 0.09% of the annual MPD. Extrapolating the data as an expression of averaged annual operator exposure resulted in a whole body dose of 0.4536 mSv (45.36 mrem) or 0.9% of the annual MPD, These results are well below established occupation exposure limits and are compatible with those published by the manufacturer.
Keywords:forensic science  radiation exposure  radiation safety  portable radiation emitting device  maximum permissive dose
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号