首页 | 本学科首页   官方微博 | 高级检索  
     检索      


The comparison of toluene determination between headspace-solid phase microextraction and headspace methods in glue-sniffer's blood and urine samples
Authors:Kim N Y  Park S W
Institution:Department of Forensic Science, National Institute of Scientific Investigation, Seoul, Korea.
Abstract:An accurate and simple method was developed to determine the level of toluene in urine and blood quantitatively by using the gas chromatography/mass spectrometry (GC/MS) with headspace--solid phase microextraction (HS-SPME) technique. An assembly of SPME with a replaceable extraction fiber, coated with 100 microm polydimethylsiloxane, was used. The detection limit of toluene in blood and urine with HS-SPME technique was 10 times higher than that with headspace (HS) technique. To compare the HS-SPME with HS technique for the determination of toluene in biological fluids, blood and urine samples from glue sniffers were analyzed by both methods. The level of toluene by the two techniques was highly correlated: the correlation coefficient (r2) between the two sets of values were 0.98 and 0.96 in urine and blood, respectively.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号