Analyse von heroinproben mit der Kapillargaschromatographie. Vergleich von Glaskapillartrennsäule und gepackter Säule |
| |
Authors: | Manfred Gloger Helmut Neumann |
| |
Institution: | Bundeskriminalamt, Kriminaltechnisches Institut/Fachgruppe Chemie, Postfach 18 20, D 6200 Wiesbaden Bundesrepublik Deutschland |
| |
Abstract: | In forensic examination of illicit heroin samples high-resolution capillary gas chromatography is compared with gas chromatography with packed columns. Separation efficiency, reproducibility of qualitative and quantitative determinations and stability of the columns are examined.A practical method with silylation of the samples is presented which allows safe identification and quantification of many various compounds contained in illicit heroin samples. The gas chromatographic data of 35 substances (opium alkaloids and synthetic derivatives, adulterants and diluents) are listed. The method also allows the quantification of morphine, opium, and cocaine samples. |
| |
Keywords: | Forensic analysis Heroin Quantification Capillary GC Column comparison |