首页 | 本学科首页   官方微博 | 高级检索  
     


Organic explosives analysis using on column-ion trap EI/NICI GC-MS with an external source
Authors:Calderara Stéphane  Gardebas Dominique  Martinez Fabienne  Khong Seu-Ping
Affiliation:Explosives department of the Institut de Recherche Criminelle de la Gendarmerie 1, Bd Theophile SUEUR-93111 Rosny Sous Bois (Paris-France).
Abstract:In this study, a standard method by gas chromatography coupled with mass spectrometry (GC/MS) for the analysis of classical organic explosives was developed. This method was validated in the EI mode, based on the XPT 90-210 standard method. Detection limits (LOD) and quantitation limits (LOQ) were both determined using electronic impact (EI) and negative ion chemical ionization (NICI) modes. These were compared and results showed that in the NICI mode, detection limits were lower than in the EI mode, thus NICI mode appeared to be the best way to analyze nitrate esters. Results of ion trap MS detection were then compared with those obtained in a previous study with single quadrupolar technology. Major ions that were obtained using ion trap MS detection in these two modes were reported.
Keywords:
本文献已被 PubMed 等数据库收录!