首页 | 本学科首页   官方微博 | 高级检索  
     


Evaluation of breath alcohol instruments II. In vivo experiments with Alcolmeter Pocket Model
Authors:A.W. Jones
Affiliation:Department of Alcohol Research, Karolinska Institutet, Stockholm Sweden
Abstract:The precision and accuracy of an Alcolmeter Pocket Model breath alcohol instrument have been investigated in experiments with human subjects under controlled conditions. The instrument response was zero in all tests with breath samples from alcohol-free subjects. The standard deviations of ethanol determinations in breath were ±0.0722 mg/ml during ethanol absorption and ±0.0416 mg/ml during ethanol elimination. The standard deviation during the elimination phase increased with ethanol concentration in the sample, being ±0.0416 mg/ml on average at a mean concentration of 0.420 mg/ml, corresponding to a coefficient of variation of 9.9%.The blood alcohol estimates using the Alcolmeter were somewhat too high during active absorption of ethanol, and too low during elimination, when a constant blood-breath alcohol ratio of 2100:1 was used to calibrate the instrument. During the elimination phase of ethanol kinetics and at a mean blood alcohol concentration of 0.50 mg/ml, the mean Alcolmeter result was 0.456 ± 0.169 mg/ml with 95% confidence, i.e. varying between 0.287 and 0.625 mg/ml 95 times out of 100 tests at this critical blood alcohol level.
Keywords:
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号