首页 | 本学科首页   官方微博 | 高级检索  
     


A micro-chemiluminescence determination of cyanide in whole blood
Authors:Lv Jiagen  Zhang Zhujun  Li Jindong  Luo Lirong
Affiliation:College of Chemistry and Material Science, Shaanxi Normal University, Xian 710062, China. lvjiagen413@yahoo.com.cn
Abstract:A reactant volume self-controlled micro-device was presented and applied to the flow injection chemiluminescence (CL) for determination of cyanide in whole blood. A mini distiller was fabricated for cyanide extraction from the blood samples with the extraction efficiencies of cyanide > or = 98%. A fluidic control platform with air driving was fabricated. The described system showed the features of easy fabrication, undiluted sample injection, safe analysis operation, and suitability for automatic cyanide analysis. The calibration curve showed linearity in the cyanide concentration range of 5.0 x 10(-7) to 5.0 x 10(-5) mol l(-1) with the detection limits (3sigma) of 2.3 x 10(-7) mol l(-1). CL peak-height precision was 1.9% R.S.D. (n = 11) at the 1.0 x 10(-6) mol l(-1) cyanide level. The new devices were applied to the analysis of cyanide in rabbit whole blood samples and the results agreed well with those obtained from official method.
Keywords:
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号