首页 | 本学科首页   官方微博 | 高级检索  
     


Strut fracture in a Bjork-Shiley aortic valve prosthesis
Authors:B D Lifschultz  E R Donoghue
Abstract:Strut fracture can be a life-threatening adverse effect of mechanical prosthetic heart valves. This complication has occurred in the DeBakey, the Beall, the Cooley-Cutter and, most recently, the Bjork-Shiley valves. We report the case of a 35-year-old man who died suddenly 16 months after a 60 degree Bjork-Shiley Convexo-Concave heart valve prosthesis was inserted in the aortic position. At autopsy, the two welded attachments of the valve's outlet strut had fractured. The valve's tilting disc was found in his abdominal aorta.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏